Quality-control a raw IFCB sample (hdr/adc/roi triplet)
Source:R/ifcb_qc_sample.R
ifcb_qc_sample.RdValidates the integrity and self-consistency of one or more raw IFCB samples
and returns a tidy tibble of QC metrics and flags, one row per sample. Each
sample is expected to consist of the standard IFCB file triplet sharing a
base name: a header (.hdr), an ADC table (.adc), and the raw image data
(.roi).
Usage
ifcb_qc_sample(
sample,
data_folder = NULL,
max_ml = NULL,
volume_tolerance = 0.05,
runtime_tolerance = 0.02,
max_roi_mb = NULL,
max_humidity = NULL,
max_temperature = NULL,
flowrate = 0.25
)Arguments
- sample
Sample(s) to check. Either a single directory (all
.adcfiles within are discovered recursively), or a character vector of sample base names or paths (with or without a.hdr/.adc/.roiextension). Whendata_folderis supplied, bare sample names are resolved against it.- data_folder
Optional directory in which to locate the triplet files when
sampleis given as bare sample names. Searched recursively.- max_ml
Optional fixed upper bound (in millilitres) for a plausible analyzed volume, applied to every sample. Default
NULLderives the ceiling per sample from the header syringe volume (SyringeSampleVolume, falling back tosyringeSize, then the 5 mL IFCB standard) scaled byvolume_tolerance. Set this only to override that instrument-reported value.- volume_tolerance
Fractional tolerance added to the derived syringe volume ceiling (default
0.05, i.e. 5%) to absorb estimation noise in the analyzed volume. Ignored whenmax_mlis supplied.- runtime_tolerance
Fractional slack by which the ADC's last-trigger run time may exceed the header's total run time before
runtime_consistentis set toFALSE(default0.02, i.e. 2%).- max_roi_mb
Optional numeric upper bound (in megabytes, where 1 MB = 1024^2 bytes) for the
.roifile size. When supplied, samples whose.roiexceeds this size are flagged in the advisoryroi_oversizedcolumn (e.g.max_roi_mb = 5for 5 MB). DefaultNULLdisables the check (roi_oversizedisNA).- max_humidity
Optional numeric threshold (percent) for the header's recorded
humidity. Samples above it are flagged in the advisoryhumidity_highcolumn. DefaultNULLdisables the check.- max_temperature
Optional numeric threshold (degrees, as recorded by the instrument) for the header's
temperature. Samples above it are flagged in the advisorytemperature_highcolumn. DefaultNULLdisables the check.- flowrate
Syringe flow rate (millilitres per minute) passed to
ifcb_volume_analyzed(). Default0.25.
Value
A tibble with one row per sample containing the resolved file paths,
QC metrics, boolean QC flags, and an overall qc_pass column.
Details
The checks build directly on the package's existing readers
(ifcb_read_hdr_data(), read_adc_columns(), ifcb_volume_analyzed())
cover four areas:
- Triplet completeness
Whether all of
.hdr,.adc, and.roiare present (files_complete).- ROI count consistency
The number of imaged ROIs in the ADC (rows with a non-zero ROI width) must equal the header's
roiCount(roi_count_match). Note that the ADC row count is not compared totriggerCount: depending on the ADC format a single trigger may yield several ROIs, sotriggerCountis reported but not used as a hard check.- ROI data completeness
The
.roifile must be at least as large as the last image's end offset (max(StartByte + width * height)) computed from the ADC. A smaller file indicates a truncated or aborted transfer (roi_data_complete).- Run time consistency
The run time recorded at the ADC's last trigger must not exceed the header's total run time (within
runtime_tolerance), since a trigger cannot fire after acquisition has stopped (runtime_consistent). A header run time materially shorter than the ADC's last trigger points to corrupted or truncated header metadata; a run that legitimately continued past the last trigger (header run time longer than the ADC's) is normal for sparse samples and is not flagged.- Flow / volume sanity
The analyzed volume from
ifcb_volume_analyzed()must be positive and not exceed the syringe sample volume (volume_ok). The ceiling is taken from the header'sSyringeSampleVolume(reported assyringe_ml, e.g. 5 mL for a standard IFCB), plusvolume_tolerance, since the analyzed volume can never physically exceed the drawn syringe volume. A fixed ceiling can be forced withmax_ml.
Further advisory flags are reported but do not affect qc_pass, as they
describe valid samples that a user may nonetheless wish to exclude:
is_bead_run (a bead/calibration run, detected from the header's runBeads
field or a sampleType containing "bead"), is_empty (no imaged ROIs),
roi_oversized (the .roi file exceeds max_roi_mb, useful for catching
overloaded or anomalous runs), and humidity_high / temperature_high (the
header's recorded humidity / temperature exceed max_humidity /
max_temperature, flagging possible condensation or overheating). The latter
three are only evaluated when their threshold is supplied; otherwise they are
NA. The measured humidity and temperature are always reported.
qc_pass is the conjunction of the integrity checks above
(files_complete, roi_count_match, roi_data_complete,
runtime_consistent, volume_ok). A check that cannot be evaluated for a
given sample is reported as NA and treated as not applicable: it does not
fail qc_pass. This matters for legacy IFCB headers, which omit the
post-run roiCount summary field (so roi_count_match is NA); such samples
can still pass on the checks that do apply. Only a check that actually
evaluates to FALSE fails the sample. files_complete and volume_ok are
always TRUE/FALSE (never NA) and so always count.
References
Sosik, H. M. and Olson, R. J. (2007), Automated taxonomic classification of phytoplankton sampled with imaging-in-flow cytometry. Limnol. Oceanogr: Methods 5, 204-216.
Examples
if (FALSE) { # \dontrun{
# Check every sample in a data directory
qc <- ifcb_qc_sample("data/raw")
# Keep only clean, non-bead samples for analysis
dplyr::filter(qc, qc_pass, !is_bead_run)
} # }